ADVERTISEMENT

Issue Navigator

Volume 07 No. 04
Earn CME
Accepted Papers
Classifieds







Scientific Investigations

Evaluation of a Single-Channel Portable Monitor for the Diagnosis of Obstructive Sleep Apnea

http://dx.doi.org/10.5664/jcsm.1196

Burcu Oktay, M.D.1; Thomas B. Rice, M.D., M.S.1; Charles W. Atwood, M.D., F.A.A.S.M.1; Michael Passero, M.D.1; Neeraj Gupta, M.D.1; Rachel Givelber, M.D., F.A.A.S.M.1; Oliver J. Drumheller, Ed.D., RRT1; Patricia Houck, M.S.H.2; Nancy Gordon, M.S.3; Patrick J. Strollo, M.D., F.A.A.S.M.1
1UPMC Sleep Medicine Center, Division of Pulmonary, Allergy, and Critical Care Medicine, University of Pittsburgh School of Medicine, Pittsburgh, PA; 2Departments of Psychiatry, Western Psychiatric Institute and Clinic, University of Pittsburgh School of Medicine, Pittsburgh, PA; 3Gordon and Associates, Berkley, CA

Study Objective:

To validate the ApneaLINK (AL) as an accurate tool for determining the presence of obstructive sleep apnea (OSA) in an at-risk sleep clinic population in a home test environment.

Methods:

Consecutive participants referred with the suspicion of OSA were evaluated in the home with the AL portable monitor (AL Home), followed by simultaneous data collection with diagnostic polysomnography (PSG) and AL in the sleep laboratory (AL Lab). Prevalence, sensitivity, specificity, and ROC curves were calculated for PSG vs. AL Lab, PSG vs. AL Home, and AL Lab vs. AL Home test. Pearson correlations and Bland-Altman plots were constructed.

Results:

Fifty-three (55% female) participants completed the entire study. The mean age of the population was 45.1 ± 11.3 years, and body mass index was 35.9 ± 9.1 kg/m2. The prevalence of an apnea hypopnea index (AHI) ≥ 15 in the cohort was 35.9%. The results demonstrated a high sensitivity and specificity of the AL respiratory disturbance index (RDI-AL) compared with the AHI from the PSG. The AL Lab had the highest sensitivity and specificity at RDI-AL values ≥ 20 events/h (sensitivity 100%, specificity 92.5%). The AL Home was most sensitive and specific at an RDI-AL ≥ 20 events/h (sensitivity 76.9%, specificity 92.5%). The Pearson correlations for PSG vs. AL Lab and PSG vs. AL Home were ρ = 0.88 and ρ = 0.82, respectively. The Bland-Altman Plots demonstrated good agreement between the methodologies.

Conclusion:

The AL home test is an accurate alternative to PSG in sleep clinic populations at risk for moderate and severe OSA.

Trial Registration:

clinicaltrials.gov ID: NCT00354614.

Citation:

Oktay B; Rice TB; Atwood CW; Passero M; Gupta N; Givelber R; Drumheller OJ; Houck P; Gordon N; Strollo PJ. Evaluation of a single-channel portable monitor for the diagnosis of obstructive sleep apnea. J Clin Sleep Med 2011;7(4):384-390.




Please login to continue reading the full article

Subscribers to JCSM get full access to current and past issues of the JCSM.

Login to JCSM

Not a subscriber?

Join the American Academy of Sleep Medicine and receive a subscription to JCSM with your membership